MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1114
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/13/2019
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A nurse reported that the a1114 mayfield infinity skull clamp was not maintaining position on (b)(6) 2019.Additional information received on 25sep2019 indicated that the pressure of 70 had dropped down to 40 during a posterior c2-c3 laminectomy for resection of neurofibroma and fusion with instrumentation bone grafting procedure.The product problem was discovered while the surgeon was trying to position the patient for the procedure.There was no patient injury reported.There was a delay in surgery due to the number of times the patient had to be repositioned in order for the skull clamp to maintain a torque pressure of 70.There was no known adverse consequence due to the delay.Additional information has been requested.
 
Manufacturer Narrative
Additional information - d10, g4, g7, h2, h3, h4, h6, h10 with respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement, when unit is properly positioned and put under pressure unit will function properly.The dhr was reviewed showed no abnormalities related to the reported failure.General maintenance and cleaning required.Unit needs heli-coils added to large starburst threads.Ratchet extension is having difficulty going in and out of the base casting due to having dimples on it.The dimples will need to be filed off.The complaint is likely caused by wear and tear / improper handling.The definite root cause cannot be reliably determined.Device identifier # (b)(4).Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(4) office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key9182554
MDR Text Key189769785
Report Number3004608878-2019-01047
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2019
Date Manufacturer Received12/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-