• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74123152
Device Problems Corroded (1131); Loose or Intermittent Connection (1371); Unstable Capture Threshold (3269); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Toxicity (2333); Injury (2348); Hip Fracture (2349)
Event Date 09/22/2017
Event Type  Injury  
Event Description
It was reported that a revision surgery on the left hip was performed due to metallosis, corrosion, pain, instability, loosening of femoral implant.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the head was removed.The cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.No further actions are required at this time.The available medical documents were reviewed.It is unknown if the underanteverted acetabular component led to accelerated wear and the debris, corrosion and synovitis noted intraoperatively.Although the pain, corrosion, synovitis and loosening of femoral implant may be findings consistent with metallosis the root cause of the reported metallosis and periprosthetic fracture cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a malperformance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK   CV31 3HL
Manufacturer Contact
sarah freestone
MDR Report Key9184021
MDR Text Key162472837
Report Number3005975929-2019-00369
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Expiration Date06/30/2016
Device Catalogue Number74123152
Device Lot Number11EW13311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/01/2019
Initial Date FDA Received10/14/2019
Supplement Dates Manufacturer Received06/17/2020
08/18/2020
01/06/2022
Supplement Dates FDA Received06/20/2020
10/22/2020
01/10/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
74122158 ACETABULAR CUP HAP SIZE 52/58 11EW06356
Patient Outcome(s) Required Intervention;