• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 201-90411
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Power Problem (3010); Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/24/2019
Event Type  Injury  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.Patient's centrimag motor is also reported under mfr# 2916596-2019-04668.
 
Event Description
It was reported that the centrimag console display screen started blinking.Flows decreased and the console display went blank.It was unsure if the flows stopped.The pump restarted but was making a grinding sound.The patient was switched to a second centrimag system.No complications.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the console display going blank and decreased and stopped flow was not confirmed.The centrimag 2nd generation primary console was not returned for analysis.No additional information or log files were provided for review.The root cause for the reported event was not conclusively determined through this analysis.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key9186123
MDR Text Key162510381
Report Number2916596-2019-04669
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-90411
Device Catalogue Number201-90411
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/24/2019
Initial Date FDA Received10/14/2019
Supplement Dates Manufacturer Received01/24/2020
Supplement Dates FDA Received02/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age47 YR
Patient Weight75
-
-