Model Number 2420-0007 |
Device Problem
Stretched (1601)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that the tubing ballooned at the silicone segment which was discovered when the pump door was opened.There was delay in care.No patient harm was reported by the customer.
|
|
Manufacturer Narrative
|
The customer¿s report that the tubing ballooned in the pump segment was confirmed.The set was visually inspected for kinks, incomplete bonding engagements, holes/tears in the tubing or damages to the components visual inspection of the set noted no damage or any anomalies.Functional testing resulted in no ballooning.Pressure testing confirmed ballooning.No issues were noted by the device.The silicone segment did not balloon.The balloon is caused by excessive pressure within the silicone segment.The root cause for the source of the excessive pressure is unknown.
|
|
Event Description
|
It was reported that the tubing ballooned at the silicone segment which was discovered when the pump door was opened.There was delay in care.No patient harm was reported by the customer.
|
|
Search Alerts/Recalls
|