Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical products: catalog# 010000996 g7 screw 6.5mm x15mm lot# 6596540.Catalog# 51-106160 tprlc 133 mp type1 pps so 16.0 lot# 6594554.Catalog# 51-108050 tprlc 133 mp t1 pps so 5x95mm lot# 6589364.Catalog# 51-145120 tprlc xr mp t1 pps 12x109mm lot# 6561974.Report source: foreign: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-04636, 0001825034-2019-04637, 0001825034-2019-04640, 0001825034-2019-04648.
 
Event Description
It was reported that debris in the sterile packaging was observed during incoming inspection.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Udi#: (b)(4).The event was confirmed with product received.Upon visual inspection, foreign debris was found inside the returned devices.Dhr was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the devices with foreign debris (not related to packaging material) when they left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.A corrective action for the foreign debris was opened to address the issue of complaints for debris in sterile packaging.The corrective action closed successfully, following the manufacture of the lot in the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 15MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9197830
MDR Text Key199909591
Report Number0001825034-2019-04635
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000996
Device Lot Number6596540
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/24/2019
Initial Date FDA Received10/16/2019
Supplement Dates Manufacturer Received02/20/2020
Supplement Dates FDA Received02/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-