Model Number M00542251
Device Problem Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the third band was attempted to be deployed; however, the band would not continue to deploy.Reportedly, the physician removed the device from the patient and the procedure was completed with another speedband superview super 7 device.Additionally, it was noted that there was no difficulty when setting up the device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the third band was attempted to be deployed; however, the band would not continue to deploy.Reportedly, the physician removed the device from the patient and the procedure was completed with another speedband superview super 7 device.Additionally, it was noted that there was no difficulty when setting up the device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Initial reporter address 1: (b)(6).Problem code 2610 for the reportable issue of bands failed to deploy.Investigation results: only the ligator head was returned for analysis.A visual examination of the ligator head found three bands present.It was also noticed that the ligator head were bent.Based on the evaluation of the returned complaint device, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.It is most likely that the ligator head teeth were damaged due to handling and manipulation of the device during the procedure.Once the ligator head teeth are damaged, the suture can be detached from its position on the ligator head and this condition could have impacted the bands deployment activity and which could have contributed to the reported issues.This failure is likely due to factors or conditions related to procedure during the use of the device that could have affected its performance and its intended purpose.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9199145
MDR Text Key177549630
Report Number3005099803-2019-05073
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/08/2020
Device Model NumberM00542251
Device Catalogue Number4225
Device Lot Number0023615811
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Date Manufacturer Received11/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age6 MO
Patient Weight64
-
-