Model Number 42544
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/06/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the pre-operative test was found to be abnormal.It was stated that the water could not enter.
 
Manufacturer Narrative
The returned valve was patent.The valve met the requirements for reflux and preimplantation testing.The valve did not meet the requirements for leak and pressure/flow testing.Water was able to flow through the valve during testing.Therefore, the nature of the complaint could not be replicated by laboratory personnel.Damage was observed to the outlet connector.It is unknown how this damage occurred.The instruction for use cautions, ¿improper handling or use of instruments when implanting shunt products may result in the cutting, slitting, crushing or breaking of components.Such damage may lead to a loss of shunt integrity¿¿ medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, 16MM, MEDIUM PRESSURE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key9199545
MDR Text Key167837520
Report Number2021898-2019-00364
Device Sequence Number1
Product Code JXG
UDI-Device Identifier00643169466913
UDI-Public00643169466913
Combination Product (y/n)N
PMA/PMN Number
K831678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number42544
Device Catalogue Number42544
Device Lot NumberE49377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2019
Date Manufacturer Received11/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-