Model Number C315HIS02 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
Blood Loss (2597)
|
Event Date 09/29/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the delivery catheter was returned and analyzed.No anomalies were found.The mechanical operation of the catheter shaft was kinked.The mechanical operation of the catheter slitting showed a spiral slit.The catheter did not slit along the score lines.There was an issue with the hemostasis valve.Visual analysis indicated damage during use.The analyst noted that a cut is visible on the hemostasis valve.It was cut at the time of slitting the catheter.Otherwise, no anomalies were found on the hemostasis valve.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the procedure, the patient experienced a lot of bleeding when the guiding catheter was inserted.The hemostasis valve was supposed to stop the bleeding.The bleeding continued through the procedure but the catheter was used as it was.It was suspected that the hemostasis valve was loose on the catheter.The catheter was removed.No further patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|