A field service engineer (fse) was at the customer site to address the reported issue.The fse confirmed the error on the error log and was able to reproduce the issue.The fse adjusted the bf probe overflow sensor and performed daily check and quality control (qc).Results were within the acceptable ranges.The aia-900 analyzer returned to operation.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4), from (b)(6) 2018 to aware date (b)(6) 2019.No other similar complaints were identified during the search period.The aia-900 operator's manual states the following: [2235] bf probe 1 suction failure cause: the overflow sensor 1 s132 detected liquid after a washer suction operation.A wu flag will be attached to the measurement result.Action: clean the wash probe 1.Check s132, the waste liquid solenoid valve sv170, the waste liquid tube, and the liquid pump lp172.The probable cause of the issue is attributed to a faulty bf probe sensor.
|
A customer reported receiving error message 2235 bf probe 1 suction failure while operating on the aia-900 analyzer.The customer recalibrated but the issue persisted.The customer verified the bf probes and replaced the probe tips, but the error remained.A field service engineer (fse) was dispatched to address the reported issue, which resulted in delayed reporting of estradiol (e2) and beta-human chorionic gonadotropin (bhcg) patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|