Model Number MSERIES BI-PHASIC |
Device Problem
Pacing Problem (1439)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device was unable to pace.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the malfunction was duplicated and the front panel membrane was replaced to remedy the problem.The customer declined repair of the device, the device was returned to the customer labeled not for clinical use.Analysis for reports of this type has not identified an increase in trend.
|
|