DEPUY SYNTHES PRODUCTS LLC BATTERY RECIPROCATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 530.715 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi: (b)(4).The reporter¿s complete address was not provided.The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the trigger of the battery reciprocator device did not move smoothly.The reporter further clarified that something got caught in the trigger when it was depressed.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the parts of the trigger did not move smoothly.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to component failure (faulty parts), which is normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|