• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNK BROACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNK BROACH Back to Search Results
Catalog Number UNK BROACH
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Literature article entitled: ¿literature received entitled: " direct, cementless, metaphyseal fixation in knee revision arthroplasty with sleeves-short-term results" written by heiko graichen, md et al., published: the journal of arthroplasty 30(2015) 2256¿2259, was reviewed for mdr reportability on 10/01/2019.The purpose of the article was to analyze the short- and mid-term results of sleeve-fixation in a large revision tka series.In the prospective study 121 patients with 193 sleeves (119 tibial and 74 femoral).In all cases the depuy pfc sigma mobile bearing revision tibial tray was used.In 77 patients, a posterior stabilized insert was used, in 27 a tc3 insert (vvc constraint), and in 17 a rotating hinge was implanted.Intraoperative complications included two fractures of the anterior, distal cortex of femur during femoral broaching.Postoperative complications included stem pain on the femoral stem tip (1.4%) and one patient with tibial stem pain a loosening of the tibial implant was noted during a later revision.4 patients have been revised to address pain and aseptic loosening.There were 14 revisions were performed.Four of those revisions were for infection.Three revision were for ligament instability, one was for malalignment and one revision was for extensor mechanism failure.There were two revisions for fracture of the implant at the junction between the stem and sleeve.Both of those cases showed insufficient osseo-integration of the sleeve in combination with malalignment and excellent stem fixation leading to low stress high cycle fatigue.The conclusion of the article noted that the cementless metaphyseal fixation with the use of sleeves is a promising option for revision tka implant fixation.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device was received.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK BROACH
Type of Device
BROACH
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6107428552
MDR Report Key9211125
MDR Text Key162974612
Report Number1818910-2019-109506
Device Sequence Number1
Product Code HTQ
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK BROACH
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/01/2019
Initial Date FDA Received10/18/2019
Supplement Dates Manufacturer Received12/19/2019
Supplement Dates FDA Received12/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-