• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNUEMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNUEMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Patient Involvement (2645)
Event Date 10/08/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver is being evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver exhibited an incorrect right pressure alarm.
 
Manufacturer Narrative
The root cause of the reported issue, incorrect right pressure alarm during routine evaluation, was determined to be a malfunction of the right pilot valve.Syncardia has a corrective and preventive action (capa) for issues related to pilot valve performance.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4998 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNUEMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9214144
MDR Text Key163063917
Report Number3003761017-2019-00303
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 01/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-