• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver is being evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
While performing a routine evaluation, a syncardia technician reported that the companion 2 driver did not pass the system check procedure.
 
Manufacturer Narrative
The reported issue of the driver not passing the system check was confirmed through the patient file review, which indicated an issue with the 9 volt battery.During investigation testing, the 9 volt battery voltage was found to be below the required range.Additional testing reproduced the results reported issue and the 9 volt battery was found to have been drained below acceptable limits.Previous investigations determined the root cause to be a malfunction of the 9 volt battery switch board assembly, causing a higher than normal current drain that results in premature 9 volt battery depletion.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 5005 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key9214146
MDR Text Key163063998
Report Number3003761017-2019-00310
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier000858000003107
UDI-Public(01)000858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number397002-001
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/14/2019
Initial Date FDA Received10/21/2019
Supplement Dates Manufacturer Received10/14/2019
Supplement Dates FDA Received11/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-