SANMINA -SCI SYSTEMS OARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number BI70002000
Device Problems Device Damaged Prior to Use (2284); Positioning Problem (3009)
Patient Problem No Patient Involvement (2645)
Event Date 10/14/2019
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the door would not close properly.There was a door open message.It was found that the door stop adjustment was out of specification from the factory.The dingus gear would not align.The door stops were properly adjusted and the alignment was then correct.The system then passed a system checkout.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that while uncrating the system from the factory, the covers applied resistance to closing the gantry and the door was difficult to close.There was no patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9214237
MDR Text Key173027350
Report Number3006544299-2019-00232
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public00763000132378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/14/2019
Initial Date FDA Received10/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-