Device evaluated by mfr: returned product consisted of an eluvia self-expanding stent system with an introducer sheath.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that pull rack was fully extended and the stent was no longer inside the device.The inner liner was separated, and the tip of the device was missing.The stent appeared to be inside the introducer sheath.The introducer sheath was x-rayed to find the missing pieces.The stent and tip of the device could be seen inside the sheath.The proximal radiopaque marker could not be seen, so the introducer sheath was cut open to verify all the pieces inside.The stent was stretched and separated in two pieces.The distal section was 4.3cm long and the proximal section was 7.5cm long.The proximal radiopaque marker was not found which indicates that part of the stent is missing.The inner liner was separated 6.8cm from the tip.There were scratch marks all over the tip of the device.There were multiple kinks along the inner liner.There was a kink to the outer sheath at the nosecone.Microscopic examination revealed no additional damages.The thumbwheel was functionally tested, and it was able to rotate freely.There is blood present inside the device.Inspection of the remainder of the device, revealed no other damage or irregularities.
|