Evaluation summary: one device was received for evaluation.The returned sample did not meet specification as received.The visual inspection found the unit arrived in fair condition, the bag deflector was scratched and the pump tray was bent.The reported condition was not duplicated.The investigation found the deficit test was conducted 3 times for two minutes each and the results were within the tolerance level.The investigation could not confirm the reported condition.An additional failure was found during the investigation; the incoming scales tests were completely off.The investigation does not indicate if the additional failure may have caused or contributed to the reported condition.The investigation did not determine a root cause or a probable root cause for the customer's report based on the information provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|