MITG - OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 7210164S
Device Problem Incorrect Measurement (1383)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a hysteroscopy, the device had a fluid deficit wherein it was measuring a negative number on the scale.It started at -300 then to -500, and then stopped at -999.The circulating nurse did a manual count of fluid deficit and was fine.There was no fluid deficit.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned sample did not meet specification as received.The visual inspection found the bag deflector damaged and there were several scratches on the design foil and display housing.The reported condition was confirmed.The investigation found the mai pole on the scale is very loose and the root cause of the failure.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG - OKLAHOMA CITY
75 s. meridian av
oklahoma OK 73107
Manufacturer (Section G)
MITG - OKLAHOMA CITY
75 s. meridian av
oklahoma OK 73107
Manufacturer Contact
lisa hernandez
5920 longbow drive
boulder, CO 80301
2034925563
MDR Report Key9216411
MDR Text Key163210297
Report Number1643264-2019-20024
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7210164S
Device Catalogue Number7210164S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-