Device evaluated by mfr: returned product consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage and the functional testing of the device was completed.The device showed a kink located 133.5cm from the tip.Functional analysis was done by completing the setup procedure per the directions for use (dfu).Aspiration testing of the device was done per the test procedure.Test results showed that this device did perform as designed per the test procedure specification sheet, withdrawing 30ml of fluid in the 1minute time frame.Inspection of the remainder of the device, apart from the observed damage revealed no damage or irregularities.
|