• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI14
Device Problems Disconnection (1171); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that inflation valve came off when the user injected water into the valve on the day of use.
 
Manufacturer Narrative
The reported event was confirmed as cause unknown.The sample was evaluated and the cap was found detached from the funnel.The exterior of the sample was inspected and no abnormalities were found.The edges of the inflation funnel were smooth and regular.The cap was manually assembled into the funnel and it was tightly attached.A potential root cause for this failure mode could funnel thickness incorrectly sized at build-up or finish dipping operations/ machine misalignment during valve/cap assembly/user related (rough handling or incorrect syringe used).The exact cause on how and when the event occurred could not be determine.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿3) using aseptic technique, position the needle-less syringe(slip-tip or luer-lock type) in the center of sampling port.The syringe should be held perpendicular to the surface of the sampling port.Press the syringe and twist to lock the syringe onto the sampling port attached luer tip syringe to the inflation valve.Inject an additional amount of sterile water into the inflation lumen and pump the plunger.If situation wouldn't be improved, serve the inflation funnel of valve." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that inflation valve came off when the user injected water into the valve on the day of use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9220409
MDR Text Key168877688
Report Number1018233-2019-06647
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016745
UDI-Public(01)00801741016745
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 11/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2022
Device Model Number0165SI14
Device Catalogue Number0165SI14
Device Lot NumberMYDN1457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2019
Date Manufacturer Received11/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-