ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A health professional reported that during an intraocular lens (iol) implant surgery, two (2) haptics broke upon placement.Additional information has been requested.
 
Manufacturer Narrative
The product was not returned.Company complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The customer indicated the use of a non-qualified company lens, 5.0 diopter lens and non-qualified company iii handpiece.The company model is only qualified for use in the company (a) cartridge.The root cause is most likely related to a failure to follow the dfu.The account used an unqualified lens/company cartridge combination.The use of non-qualified combinations may result in delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9225346
MDR Text Key170310532
Report Number1119421-2019-01773
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/16/2019
Initial Date FDA Received10/23/2019
Supplement Dates Manufacturer Received12/13/2019
Supplement Dates FDA Received12/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MN60MA IOL; MONARCH III HANDPIECE
-
-