• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL TECHNOLOGY, INC. AMT BRIDLE PRO; NASAL TUBE RETAINING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL TECHNOLOGY, INC. AMT BRIDLE PRO; NASAL TUBE RETAINING SYSTEM Back to Search Results
Model Number 4-42XX
Device Problems Inadequate or Insufficient Training (1643); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
This report is a response to user facility report # (b)(4).At the time of the report the device was still in use and had not been removed from the patient.Based on the provided information the reported incident was not caused by a manufacturing defect.The device functioned as intended and there was no malfunction as defined by the fda.Per the bridle directions for use once the bridle has been placed the excess tubing and magnet is intended to be cut and disposed of.Based on the provided information this was not done prior to the mri scan being initiated.The bridle device is mr safe when placed according to the instructions for use, which the user did not properly follow.We have assigned complaint # (b)(4) to this report and will provide additional information to the fda if the device or additional information is able to be obtained and its analysis changes the conclusion of this report.
 
Event Description
While moving the patient into the scan room, the rad tech assistant noticed that the end of the patient's feeding tube moved as they got closer to the magnet.The patient was removed from the scan room and it was discovered that there was a small, black, piece of metal on the tip of the feeding tube (on the end that was hanging out of the patient's nose).Prior to entering the scan room, the patient had been wanded and this tip did not alarm.The mri technologist reviewed the patient's chart and found that the patient had an amt bridle nasal tube.Per the manufacturer's website, the apparatus was "mr safe".Because metal is never considered "mr safe" more research was conducted.It was discovered that the metal tip was a magnet that should have been removed after the nasal tube was placed.The clinical administrator was called to mri.Clinical administrator stated that the magnet should have been clipped off and he did this himself with a pair of scissors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMT BRIDLE PRO
Type of Device
NASAL TUBE RETAINING SYSTEM
Manufacturer (Section D)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine blvd.
brecksville OH 44141
Manufacturer (Section G)
APPLIED MEDICAL TECHNOLOGY, INC.
8006 katherine blvd.
brecksville OH 44141
Manufacturer Contact
sarah beercheck
8006 katherine blvd.
brecksville, OH 44141
4402622541
MDR Report Key9226704
MDR Text Key210633163
Report Number1526012-2019-00007
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number4-42XX
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/23/2019
Initial Date FDA Received10/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-