The initial reporter stated they received discrepant results for one patient sample tested with the elecsys ft4 iii assay on a cobas 8000 e 801 module.It was asked, but it is not known if any incorrect results were reported outside of the laboratory.Refer to the attachment for all patient data.The sample was initially tested on the customer's e 801 analyzer on (b)(6) 2019.The sample was repeated on a centaur analyzer.The sample was also provided for investigation, where it was tested on a second e 801 analyzer on (b)(6) 2019, a cobas e 411 immunoassay analyzer on (b)(6) 2019, and a cobas 8000 e 602 module on (b)(6) 2019.The serial number of the customer's e 801 analyzer is (b)(4).The serial number of the e 801 analyzer used for investigation is (b)(4).Ft4 reagent lot number 380330, with an expiration date of 31-dec-2019 was used on this analyzer.The serial number of the e 411 analyzer used for investigation is (b)(4).Ft4 reagent lot number 391521, with an expiration date of 29-feb-2020 was used on this analyzer.The serial number of the e 602 analyzer used for investigation is (b)(4).Ft4 reagent lot number 391521, with an expiration date of 29-feb-2020 was used on this analyzer.
|