• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION SET Back to Search Results
Catalog Number JC-05400-B
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that when pulling up the syringe the air was entered syringe and the blue cone of the syringe came loose.
 
Manufacturer Narrative
(b)(4).A lot number was not provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the lor syringe with no relevant findings.A design history review was performed for part # kz-05501-002 as a part of this complaint investigation.Per eco-051699 (released 03-dec-2018), supplier (preox) made the following changes: kz-05501-002 luer plastic lor syringe: changed plunger material from profax 535 to profax 531.Changed to new plunger tool.Changed to new mold for blue stopper.Changed molding location for the plunger and the blue stopper as follows.Plunger: from fleimaplastic in germany to gpe, germany.Blue stopper: from et, germany to psilkon, germany.These effective changes did impact product design and material.Complaint verification testing could not be performed as no sample was provided for analysis.A lot number was not provided.A lot number was not provided.However, the customer did provide photos of the lidstock.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the lor syringe with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.
 
Event Description
It was reported that when pulling up the syringe the air was entered syringe and the blue cone of the syringe came loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key9233487
MDR Text Key170251166
Report Number3006425876-2019-00869
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJC-05400-B
Device Lot NumberUNKN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/07/2019
Initial Date FDA Received10/24/2019
Supplement Dates Manufacturer Received11/21/2019
Supplement Dates FDA Received11/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-