Model Number FT4 G3 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/24/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sample from the patient was requested for investigation.This event occurred in (b)(6).
|
|
Event Description
|
The initial reporter received questionable elecsys ft4 iii results for one patient sample from the cobas e 801 module serial number: (b)(4).The sample was repeated on a centaur analyzer and on a cobas e 801 module, a cobas e 602 module, and cobas e411 analyzer.Refer to the attachment to the medwatch for all patient data.It was not known if any erroneous result was reported outside of the laboratory.
|
|
Manufacturer Narrative
|
Sample from the patient was received for investigation.An interfering factor to streptavidin in the assay antibody/idiotype could be identified in the sample.This interference is documented in product labeling for the assay as "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.".
|
|