Device identifier #: (b)(4) the device was returned for evaluation.Device history record reviewed with no abnormalities related to the reported failure.The device passed all required inspection points with no associated mrr¿s, variances or rework.Sampling plan reviewed and is acceptable.Device is 100% tested prior to final acceptance.The reported complaint was confirmed from the evaluation of item.With respect to the complaint, it was confirmed the returned unit did not meet all specific functional test.Unit received with the mayfield 2 lock has up and down movement and the teeth are grinding when rotated.General maintenance and cleaning required.The complaint is likely caused by wear and tear / improper handling.The definite root cause cannot be reliably determined.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(4), director of regulatory programs, office of product evaluation and quality and (b)(4), assistant director, mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
|