MAUDE Adverse Event Report: AESCULAP AG IQ MEDIALIZED TIBIA CUTTING GUIDE LEFT; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG IQ MEDIALIZED TIBIA CUTTING GUIDE LEFT; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NS406R
Device Problems Material Frayed (1262); Sharp Edges (4013)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: the complained devices are available in a decontaminated condition for investigation.Tibia cutting guide has a metal burr on the side.According to the available information, there were no negative consequences for patient.Investigation: sharp-edged burrs at the outer ends of the cutting slot.The components were examined visually and microscopically with the digital microscope vhx-5000 keyence eq.-nr.2000024840 and the digital-camera "panasonic dmc tz8".The complained devices were sent to the responsible manufacturing department for investigation: "on the production side no failure could be found." batch history review: the device quality and manufacturing history records have been checked for the available lot numbers and found to be according to our specification valid at the time of production.Conclusion and root cause: based on the information available as well as a result of our investigation the root cause of the failure is most probably usage related.Rationale: there are no hints for a material problem.According to the quality standard and dhr files a material defect and production error was not found.According to the manufacturing process and specified targets it can be excluded that the device has left aag with such sharp-edged burrs at the outer ends of the cutting slot.Most probably the device was not used as intended.
 
Event Description
It was reported that there was an issue with an iq medialized tibia cutting guide.The tibia cutting guide has a metal degree on the side.This results in a risk of injury for the surgeon.Tibia cutting guide has a metal burr.There was no patient harm.Additional information was not provided nor available.The malfunction is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2019-00726 ((b)(4) ns407r).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IQ MEDIALIZED TIBIA CUTTING GUIDE LEFT
Type of Device
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
kerstin rothweiler
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key9234916
MDR Text Key176780377
Report Number9610612-2019-00727
Device Sequence Number1
Product Code HTZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNS406R
Device Catalogue NumberNS406R
Device Lot Number52516441
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-