Model Number R SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "check pads" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The reported malfunction was duplicated and attributed to a faulty fuse on the analog board.The device was rectified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|