• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74122546
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problems Failure of Implant (1924); Toxicity (2333)
Event Date 10/15/2019
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed due to elevated chrome levels.Serum 2.9 ng/mm cobalt 16.2 ng/mm.
 
Manufacturer Narrative
It was reported that revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and sleeve, was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the hemi head and bhr cup.Similar complaints have been identified for the sleeve and this failure will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as a result of the reported event.Cobalt was stated to be 16.2 ng/mm.However, no clinical/medical documentation has been provided as support for this complaint.Without supporting clinical/medical documents, a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 46MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key9245578
MDR Text Key163980074
Report Number3005975929-2019-00387
Device Sequence Number1
Product Code NXT
UDI-Device Identifier00885556071588
UDI-Public00885556071588
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/28/2012
Device Model Number74122546
Device Catalogue Number74122546
Device Lot Number07JW13859
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received10/28/2019
Supplement Dates Manufacturer Received10/15/2019
Supplement Dates FDA Received02/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ACETLR CUP HAP 54MM 74120154/080030; MODULAR SLEEVE 74222200/ 07EW11882; ACETLR CUP HAP 54MM 74120154/080030; MODULAR SLEEVE 74222200/ 07EW11882
Patient Outcome(s) Hospitalization; Required Intervention;
-
-