It was reported that revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head and sleeve, was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the hemi head and bhr cup.Similar complaints have been identified for the sleeve and this failure will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as a result of the reported event.Cobalt was stated to be 16.2 ng/mm.However, no clinical/medical documentation has been provided as support for this complaint.Without supporting clinical/medical documents, a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|