SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 5; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 74012415
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2019
Event Type  malfunction  
Event Description
It was reported a size 5 j2 5 in 1 cutting block fall apart when pinned to the patient while dr.Was making cuts through it.The central housing cane unscrewed and fell out.All pieces were retrieved from the wound.The case was delayed for 3-5 minutes while we cleaned up the pieces, changed gloves and irrigated to account for possible debris being contaminated.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The screw that retains the crossbar came out causing the device to disassemble and become inoperable.All pieces were returned.The device was manufactured in 2017 and exhibits signs of extensive wear/ usage.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY DCF AP FEM CUT BLK 5
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9251270
MDR Text Key164148166
Report Number1020279-2019-03771
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010556677
UDI-Public03596010556677
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number74012415
Device Lot Number17CM15749
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2019
Initial Date Manufacturer Received 10/02/2019
Initial Date FDA Received10/29/2019
Supplement Dates Manufacturer Received11/11/2019
Supplement Dates FDA Received11/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-