Model Number G31577
Device Problem Failure to Align (2522)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2019
Event Type  malfunction  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all captura biopsy forceps without spike are subjected to a visual inspection and functional testing to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
Prior to an esophagus endoscopy procedure, the physician uncovered a captura biopsy forceps without spike but observed the device had a defect of misalignment.This defect was evident when the forceps were dissembled [unpackaged].This observation was made prior to patient contact.
 
Manufacturer Narrative
A correction was made to the manufacturing and expiration dates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA BIOPSY FORCEPS WITHOUT SPIKE
Type of Device
PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key9252646
MDR Text Key220573553
Report Number1037905-2019-00653
Device Sequence Number1
Product Code PTS
UDI-Device Identifier00827002315778
UDI-Public(01)00827002315778(17)211001(10)W4125971
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/01/2021
Device Model NumberG31577
Device Catalogue NumberDBF-2.4-160-S
Device Lot NumberW4125971
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/02/2019
Initial Date FDA Received10/29/2019
Supplement Dates Manufacturer Received03/23/2022
Supplement Dates FDA Received04/08/2022
Date Device Manufactured10/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-