• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. HIP IMPLANT; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. HIP IMPLANT; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL Back to Search Results
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Injury (2348)
Event Date 10/10/2019
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed due to pain and migration.Explanted device: bipolar (tandem) / implanted device: femoral head (g-taper head).
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that the requested medical records have not been received on this complaint.Although an undated x-ray was provided, a conclusion cannot be made regarding the revision.Without supporting medical evidence, the reported event cannot be assessed and a thorough medical assessment cannot be performed.Upon receipt of medical/clinical records, the clinical task can be re-opened and assessed at that time.Some potential probable causes of the reported issue could include poor bone quality or improper device selection.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP IMPLANT
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9255798
MDR Text Key164290586
Report Number1020279-2019-03807
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Initial Date Manufacturer Received 10/10/2019
Initial Date FDA Received10/30/2019
Supplement Dates Manufacturer Received01/30/2020
Supplement Dates FDA Received02/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-