• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO BONE SCREW DRIVER; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO BONE SCREW DRIVER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110164
Device Problems Material Separation (1562); Solder Joint Fracture (2324)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2019
Event Type  Injury  
Event Description
It was reported that the inside of the pin driver, that supports the pin, detached from the rest of the driver while inserting bone pins.Doctor used speed pin adapter from journey set to complete procedure.No delays and no patient injuries involved.
 
Manufacturer Narrative
The device was used for treatment.The functional inspection confirmed that the inner cylinder of the pin driver would spin freely when spinning it around a bone screw.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specifications or would not be able to perform as intended.A complaint history review identified similar events.This failure mode will be trended to assess for any necessary corrective actions.This failure is an identified failure mode within the risk file.The navio surgical technique guide (pn 500099 rev c) released at the time of the complaint provides instructions on the navio instrumentation kit.It states that ¿the navio instrument kit consists of a two-level tray that contains all of the required instrumentation for surgery using the navio surgical system.If the equipment breaks or fails during surgery, a sterile backup tray is on-site and can be unwrapped to replace a broken or dropped tool.¿ it also has a warning which states:¿ a full traditional surgical instrumentation tray for the chosen implant should be available during every system use to manually implant the prosthesis in the event of system failure.¿ per complaint details, the device malfunction occurred during surgery.Based on the information provided, the procedure used the journey instrumentation set to complete the surgery.There was no reported delay, no injury/impact to the patient; therefore, no further medical assessment is warranted at this time.We were able to confirm there was a relationship established between the reported event and the device.The malfunction is due to a deficiency of the pin driver where the laser weld that attaches the outer tube to the inner cylinder with the triangular geometry fails, causing the outer tube to spin without spinning the screw.This issue is being further investigated and corrective action has been requested.
 
Manufacturer Narrative
H10: a1, d10, h3: updated information.H11: b1, b2, g5, h1, h6: corrected information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO BONE SCREW DRIVER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key9258175
MDR Text Key188489258
Report Number3010266064-2019-00127
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628614
UDI-Public00885556628614
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110164
Device Catalogue NumberPFSR110164
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2019
Initial Date Manufacturer Received 10/08/2019
Initial Date FDA Received10/30/2019
Supplement Dates Manufacturer Received04/23/2020
08/10/2020
Supplement Dates FDA Received04/27/2020
08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-