(b)(4).Finished goods records of mic4036, pebcqqq0 have been reviewed and show no deviations; all results meet the specified quality requirements.Additional information was requested and the following was obtained: package lot number of the clips? ¿ lot number was included in the complaint.Was the clips difficult to load on the device? ¿ it was difficult to load and with some even impossible to load.Were clips used in the procedure? - yes.Were you able to lock the clip closed on the suture? if yes, after it closed, was the clip.Holding securely fixed on the suture? with most it was not possible to clip and close.What suture type was used? polysorb.What suture size was used? 3-0.When the even occurred, was the suture placed near the hinge of the clip? ¿ the suture was placed normally into the clip.Was the applier checked for damage (jaws straight and aligned)? ¿ yes the applier was checked for damage and it was send to repair/ align ¿ no damage was found.Investigation summary: one mic4036 reload was received with no apparent damaged and with only 3 clips loaded.The cartridge was tested for functionality with test device.Upon functional testing of the device, the instrument loaded, retained and deployed the 3 clips as intended.The clips were form as intended and conforming to our manufacturing requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.The event described could not be confirmed as the clips performed without any difficulties noted.
|