ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problems Fracture (1260); Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/11/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided when the evaluation is completed.
 
Event Description
Related mfr report: 1627487-2019-12226.It was reported the patient's scs lead exhibited high impedance.The lead was replaced, during the procedure it was found the lead fractured at the anchor site.The anchor was found broken at the nose.Two new leads were implanted with new anchors.Postoperative, the new system was successfully programmed and stimulation therapy restored.
 
Event Description
Related mfr report: 1627487-2019-12226.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9264648
MDR Text Key164594138
Report Number1627487-2019-12227
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/05/2020
Device Model Number1192
Device Catalogue Number1192
Device Lot Number6550945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2019
Initial Date FDA Received10/31/2019
Supplement Dates Manufacturer Received11/22/2019
Supplement Dates FDA Received11/25/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-