• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problems Unintended Ejection (1234); Failure to Form Staple (2579)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t92k0d.Investigation summary: the analysis results found that the er320 device was returned for analysis and upon inspection the jaws were found to be in a yielded condition.In an attempt to replicate the reported incident the device was functionally evaluated.Upon firing of the device, the remaining 12 clips were ejected due to the condition of the jaws.Finally, the instrument locked out as intended.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.Possible causes for the found condition of the yielded jaws may be caused by the device being closed over an existing hard object or clip, placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position, or excessive application of torque to the jaws when positioning the device on a vessel.The reported complaint was confirmed.It is known from the history of the device that the condition of the jaws may lead dropping/ejected clips.A manufacturing record evaluation was performed for the finished device lot/batch number and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, each time the user fired the ligaclip, the clips came out open and where spread around into the abdomen.The clips were ejected from the jaws of the device on the tissue "open." the surgery was completed with a covidien endoclip device.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key9264720
MDR Text Key199877833
Report Number3005075853-2019-23143
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2019
Date Manufacturer Received10/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-