• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC POWER MODULE FOR TRAUMA RECON SYSTEM; BATTERY, REPLACEMENT, RECHARGEABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC POWER MODULE FOR TRAUMA RECON SYSTEM; BATTERY, REPLACEMENT, RECHARGEABLE Back to Search Results
Catalog Number 05.001.202
Device Problem Complete Loss of Power (4015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).The reporter's facility address was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the power module device stopped working.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the light-emitting diode (led) warning light indicator was not working properly and was lighting up, the device was leaking liquid and the indicator on hall sensor was red.The device also failed pretest for check indicator ¿ liquid damage, function- test, charging and checking of power module in charger and information button & self-test.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to user.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER MODULE FOR TRAUMA RECON SYSTEM
Type of Device
BATTERY, REPLACEMENT, RECHARGEABLE
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9266908
MDR Text Key195946555
Report Number8030965-2019-69886
Device Sequence Number1
Product Code MOQ
UDI-Device Identifier07611819977822
UDI-Public7611819977822
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.202
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2019
Date Manufacturer Received11/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-