• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. IMPRESS PERIPHERAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. IMPRESS PERIPHERAL CATHETER Back to Search Results
Catalog Number 510038BER
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2019
Event Type  Injury  
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The account alleges that during an endovascular aortic aneurysm repair procedure, a catheter tip detached within the patient.The physician had acquired retrograde femoral artery access.During catheter manipulations over a stiff guidewire, the catheter kinked and detached within the patient's ipsilateral common iliac artery.The physician was able to successfully remove the catheter tip from the patient with a vascular snare device.No additional patient consequences to report.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.No definitive root cause could be determined however, it is likely that significant force was applied to the device during use.The device history record was reviewed and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPRESS PERIPHERAL CATHETER
Type of Device
PERIPHERAL CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston, TX 77047
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
14646 kirby drive
houston, TX 77047
Manufacturer Contact
david lockridge
1600 merit parkway
south jordan, UT 84095
8012084551
MDR Report Key9266964
MDR Text Key164666801
Report Number3010665433-2019-00071
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00884450011485
UDI-Public00884450011485
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K053171
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date05/31/2022
Device Catalogue Number510038BER
Device Lot NumberE1600006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-