• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY; EXTERNAL POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY; EXTERNAL POWER SUPPLY Back to Search Results
Catalog Number 295400-001
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/15/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom hospital ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom hospital ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom driver hospital ac power supply had a cracked connector while supporting a patient.There was no reported adverse patient impact.
 
Manufacturer Narrative
Visual inspection confirmed the customer-reported issue as the connector outer housing was found to be cracked.It is likely that the damage to the connector outer housing was caused by an excessive amount of force being applied to the connector and outer housing.This can happen from someone stepping on the unit, something falling on the unit, or someone trying to force the unit into an adaptor from the wrong angle.Despite the physical damage, the power supply functioned as intended.It cannot be conclusively determined what caused the housing to crack, but it was likely caused by rough handling.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce (b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOSPITAL AC POWER SUPPLY
Type of Device
EXTERNAL POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az AZ 85713
MDR Report Key9267074
MDR Text Key164680212
Report Number3003761017-2019-00313
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number295400-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Initial Date Manufacturer Received 10/15/2019
Initial Date FDA Received11/01/2019
Supplement Dates Manufacturer Received10/15/2019
Supplement Dates FDA Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age28 YR
-
-