Model Number 774F75
Device Problem Material Rupture (1546)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2019
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe and two three-way stopcocks was returned for evaluation.A non-edwards contamination shield was located on the catheter body between 1.5 cm and 37 cm proximal from the catheter tip.Clotted blood was observed from the catheter.The balloon was found to be ruptured.The ruptured edges of balloon latex appeared to be different shapes and were not able to match up.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality was observed from the catheter body or returned syringe.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon issue was confirmed during the analysis.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter ruptured on the first day of use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER
Type of Device
SWAN GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key9267600
MDR Text Key188719619
Report Number2015691-2019-04070
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/06/2020
Device Model Number774F75
Device Catalogue Number774F75
Device Lot Number61719622
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2019
Initial Date Manufacturer Received 10/11/2019
Initial Date FDA Received11/01/2019
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received01/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-