• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.As received, the catheter tip was found to be bent.No visible damage or abnormality to the catheter body, balloon, windings or returned syringe was found.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 minutes without leakage.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of pacing issue could not be confirmed during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.No further actions will be taken at this time.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the swan ganz catheter was unable to pace from the beginning of use.The catheter was exchanged and the problem was solved.It is unknown if the patient had cardiac conduction defect.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key9267632
MDR Text Key188722929
Report Number2015691-2019-04071
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/20/2021
Device Model NumberPE074F5
Device Lot Number62310898
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2019
Initial Date Manufacturer Received 10/11/2019
Initial Date FDA Received11/01/2019
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received01/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-