(b)(4).Batch # r94n9a.Investigation summary: the device was returned with the blade tip broken off and returned.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device will stop activating when the blade becomes damaged.The device was disassembled to inspect the blade and the blade breaking point was located at an area inside the tube proximal to the tissue pad.This is consistent with a side load being applied to the blade while active with the jaw closed.Due to the condition of the device, the reported event of "tissue effect issues" could not be confirmed.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.No conclusion could be reached as to how this issue occurred.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|