The damaged spring in the cpc connector could not be confirmed as the cpc connector was discarded by the customer.Clinicians and patients are trained to thread a wire tie beneath the thumb release tab of the cpc connectors to prevent accidental disconnection of the cannulae from the drivelines.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its evaluation and is closing this file.(b)(4).
|
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that during a driver exchange at the hospital, it was discovered that the cpc connector on the tah-t cannula had a displaced spring.The customer also reported that the physician replaced the connector.There was no reported adverse patient impact.
|