MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD INFINITY SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1114
Device Problems Device Slipped (1584); Unintended Movement (3026)
Patient Problem Laceration(s) (1946)
Event Date 10/15/2019
Event Type  Injury  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report no.: 3004608878-2019-01074.
 
Event Description
This is 1 of 2 reports.A customer reported that there may have been an issue with the torque knob/screw, accuracy of the psi indicator and/or the locking mechanism of the a1114 mayfield infinity skull clamp.The patient underwent a craniotomy for tumor resection on (b)(6) 2019.The surgeon was doing a stealth navigation when the patient started coughing and slipped from the skull clamp.The patient incurred a laceration on the posterior of the head.The surgeon immediately applied pressure to the site and stapled the laceration.The surgeon thought at first that the patient slipped from the pins because she was bucking hard.The patient was repositioned.The skull pins were changed but the same skull clamp was used.There was a delay in surgery of about 20 to 30 minutes.
 
Manufacturer Narrative
Device identifier # (b)(4) the product was not sent in to the manufacturer for evaluation.The dhr documentation was reviewed and no anomalies that could be associated with the complaint incident were observed.The complaint was unconfirmed.The device passed all required inspection points with no associated mrr¿s, variances or rework.Sampling plan reviewed and is acceptable.Device is 100% tested prior to final acceptance.No root cause was determined.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(4), director of regulatory programs, office of product evaluation and quality and (b)(4), assistant director, mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD INFINITY SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key9286672
MDR Text Key166980634
Report Number3004608878-2019-01073
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K051440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberA1114
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/16/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received01/02/2020
Supplement Dates FDA Received07/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-