The returned valve had a catheter connected to the outlet connector with one suture.The valve met the requirements for valve flux, patency and reflux testing.The valve did not meet the requirements for leakage, pressure-flow and preimplantation testing.Tears on the to the flange, proximal occluder and reservoir were observed.It is unknown how or when this damage may have occurred.The instructions for use cautions, ¿handling or the use of instruments when implanting these products may result in the cutting, slitting, crushing, or breaking of components.Use of sharp instruments while handling these devices can nick or cut the silicone elastomer, resulting in leakage and necessitating revision.Care must also be taken when closing incisions to ensure that the devices are not cut or nicked by suture needles.Such damage may lead to loss of product integrity¿¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|