• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD S CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA¿ QUAD S CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBB1QQ
Device Problem Electromagnetic Interference (1194)
Patient Problem Complaint, Ill-Defined (2331)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 690r30 heart ring, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient had loss of cardiac resynchronization therapy due to oversensing.The implantable cardioverter defibrillator (icd) had electromagnetic interference (emi) oversensing.It was surmised that the emi was due to interference from the patient¿s left ventricular assist device (lvad).The device remains in use.It was further reported that the right ventricular (rv) lead triggered a lead integrity alert (lia) due to oversensing with high rate non-sustained (hr-ns) and sensing integrity counter (sic) episodes.The lead had t-wave oversensing (twos).The lead had diminished r-wave sensing and the pacing impedance dropped notably with an impedance out of range warning.The lead was reprogrammed and the lead issues were resolved.The lead continues to be monitored and remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing due to electromagnetic interference/noise.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ QUAD S CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9287479
MDR Text Key176769710
Report Number3004209178-2019-21191
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169007161
UDI-Public00643169007161
Combination Product (y/n)N
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2017
Device Model NumberDTBB1QQ
Device Catalogue NumberDTBB1QQ
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/10/2019
Initial Date FDA Received11/06/2019
Supplement Dates Manufacturer Received01/07/2020
Supplement Dates FDA Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6947M62 LEAD, 4298-88 LEAD, 5076-52 LEAD
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient Weight91
-
-