A field service engineer (fse) was at the customer site to address the reported issue.The fse was not able to confirm or reproduce the issue.The fse verified all adjustments and temperatures, and all were within range.A precision test was performed on calibration level 3 and mac controls; all were within ranges.The customer then analyzed new patient samples, and all results were within the patient clinical range.The aia-360 analyzer operated as expected and returned to operation.No further action required by field service.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 07sep2018 to aware date (b)(6) 2019.No other similar complaints were identified during the search period.The st aia-pack e2 analyte application manual states the following: limitations of the procedure.For diagnostic purposes, the results obtained from this assay should be used in conjunction with other data (e.G.Symptoms, results of other tests, clinical impressions, therapy, etc.).Using st aia-pack e2, the highest concentration of estradiol measurable without dilution is approximately 3000 pg/ml, and the lowest measurable concentration is 25 pg/ml (assay sensitivity).Although the approximate value of the highest calibrator is 3250 pg/ml, the exact concentration may be slightly different.The assay specification, assay range high, should be defined as the upper limit of the assay range, 3000 pg/ml.Although hemolysis has an insignificant effect on the assay, hemolyzed samples may indicate mistreatment of a specimen prior to assay and results should be interpreted with caution.Lipemia has an insignificant effect on the assay except in the case of gross lipemia where spatial interference may occur.Specimens from patients with hyperbilirubinemia may yield falsely elevated results.Certain medications may interfere with assay performance.Specimens from patients taking medicines and/or medical treatment may show erroneous results.All results should be interpreted with respect to the clinical picture of the patient.For a more complete understanding of the limitations of this procedure, please refer to the specimen collection and handling, warnings and precautions, storage and stability, and procedural notes sections in this insert sheet.Expected values.Each laboratory should determine a reference interval which corresponds to the characteristics of the population being tested.As with all diagnostic procedures, clinical results must be interpreted with regards to concomitant medications administered to the patient.The probable cause of the issue could not be determined with the available information.All related mdrs associated with this event: 8031673-2019-00429 8031673-2019-00452 8031673-2019-00453 8031673-2019-00454 8031673-2019-00455.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|