Model Number SPSAAJ |
Device Problems
Break (1069); Disconnection (1171); Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/02/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint f&p sleepstyle series cpap is currently en route to fisher & paykel healthcare in (b)(4) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
|
Event Description
|
A distributor in (b)(6) reported that upon removing the power cord from the back of the f&p sleepstyle auto cpap, the inlet pin from the power inlet socket had come apart and was stuck with the power cord.There was no reported patient consequence.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint f&p sleepstyle series cpap was received at fisher & paykel healthcare in new zealand and was visually inspected and externally.Results: visual inspection confirmed that one pin had detached from the mains inlet socket.Conclusion: the reported incident was traced to an issue in the assembly process of the supplied mains inlet connector component.The supplier of the component was notified and they have made changes to the assembly process.As part of our ongoing product improvement initiatives, we recently implemented a gauge test which identifies and rejects any potentially faulty mains inlet sockets prior to assembly into the sleepstyle.The subject sleepstyle was manufactured prior to implementation of these measures.Our user instructions that accompany the f&p sleepstyle state the following: do not use if the device, power cord or accessories are damaged, deformed, or cracked.Do not pull on the power cord as it may become damaged.Turn the device off at the power supply, then remove the power cord from the rear of the device.
|
|
Event Description
|
A distributor in japan reported that upon removing the power cord from the back of the f&p sleepstyle auto cpap, the inlet pin from the power inlet socket had come apart and was stuck with the power cord.There was no reported patient consequence.
|
|