It was reported that after 2 weeks of shunt installation, the valve was not working properly.It was stated that it stopped working with pressure 140-160, but only up to 170.The valve was explanted and replaced with another valve.The patient was implanted for triventricular hydrocephalus.
|
The returned valve was patent.The valve met the requirements for leak, siphon, and reflux.The valve did not meet the requirements for pressure-flow and preimplantation testing.Biological debris was observed on the interior of the valve.The instructions for use cautions, ¿shunt obstruction may occur in any of the components of the shunt system.The system may become internally occluded due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ all valves are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|