MAUDE Adverse Event Report: TOSOH CORPORATION AIA-360

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION AIA-360 Back to Search Results
Model Number AIA-360
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); High Readings (2459)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/14/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluation by manufacturer: the aia-360 operator's manual under section 7-2: list of flags states the following: dl is generated when the substrate dispensing amount is less than the specified amount or the lamp intensity of the fluorescence detector is low.Solution: if the dl flag is attached, replace the substrate and repeat assay.If the dl flag appears again, contact the service department.The st aia-pack intact pth application analyte manual states: evaluation of results: quality control - in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause of the reported event is pending investigation completion.
 
Event Description
A customer reported out of range high quality control (qc) result on the aia-360 analyzer.The customer stated the qc result was high for both fresh qc and 3 days old qc.Customer also reported getting dl flag on the repeat qc run.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient samples for intact parathyroid hormone (ipth).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Event Description
N/a.
 
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.The 3-way solenoid valve was returned to tosoh instrument service center for investigation.Visual inspection confirmed the wiring of the valve was cut, therefore no further testing was performed.The most probable cause of the reported event was due to faulty 3-way solenoid valve.
 
Manufacturer Narrative
Please refer to g4 for corrected data.The date received by manufacturer was missing from the submitted follow up #1.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
AIA-360
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
MDR Report Key9297705
MDR Text Key220346376
Report Number8031673-2019-00439
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 01/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360
Device Catalogue Number019945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Date Manufacturer Received11/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-