Device evaluation by manufacturer: the aia-360 operator's manual under section 7-2: list of flags states the following: dl is generated when the substrate dispensing amount is less than the specified amount or the lamp intensity of the fluorescence detector is low.Solution: if the dl flag is attached, replace the substrate and repeat assay.If the dl flag appears again, contact the service department.The st aia-pack intact pth application analyte manual states: evaluation of results: quality control - in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause of the reported event is pending investigation completion.
|